Sign in →

Test Code HGML Hemogram (CBC Without Differential)

Additional Codes

EPIC LAB4048

Methodology

Sysmex XN-2000

Sample Type

Preferred Sample Type
Lavender top - EDTA Whole Blood

Centrifuge: No

Specimen Minimum Volume

Closed tube: 1 mL

Open tube: 300 µL

Open microtube: 160 µL

Specimen Stability

Temperature Time
Ambient (18-25°C) (store) 24 hours
Refrigerated (2-8°C) 48 hours

Reference Ranges

Normal Ranges
Test 0 - 1 Month 1 - 12 Months 1 - 2 Years 2 - 10 Years >10 Years Male >10 Years Female Units
WBC 4 - 21 3.5 - 13 2 - 12 3.5 - 11.5 3.8 - 11 3.8 - 11 109/L
RBC 3.3 - 5.9 2.9 - 5.5 3.7 - 5.4 3.5 - 5.4 4.2 - 5.7 3.7 - 5.1 106/uL
HGB 10 - 21.5 9 - 15 10 - 13.5 10 - 15 13.2 - 17 11.3 - 15.5 g/dL
MCV 88 - 120 70 - 104 70 - 86 72 - 95 80 - 100 80 - 100 fL
HCT 30 - 65 26.5 - 45 30.5 - 40.5 31 - 42 39 - 50 34 - 46 %
MCH 30.5 - 42 25 - 37 20 - 34 23.5 - 34 27 - 34 27 - 34 pg
MCHC 32 - 36 32 - 36 32 - 36 32 - 36 32 - 36 32 - 36 g/dL
RDW 14 - 22 10 - 20 10 - 20 10 - 15 11 - 15.5 11 - 15.5 %
PLT 140 - 450 150 - 650 150 - 550 150 - 450 150 - 400 150 - 400 109/L
MPV 6 - 14 5.5 - 11 6 - 10 6 - 12 7 - 14 7 - 14 fL

 

Critical Values
Test Values Units
WBC <0.2 or >50.0 109/L
Hemoglobin <5.0 g/dL
Hematocrit <15.0 %
Platelets (<1 month) <20 or >998 109/L
Platelets (>1 month) <10 or >998 109/L

 

Rejection Criteria

  • Clotted specimens
  • Short draws
  • Non-EDTA samples
  • Grossly hemolyzed samples
  • Specimens outside of stability limits

Availability

Performed STAT
24/7 Yes

 

Performing Laboratory

MultiCare Yakima Memorial Hospital Laboratory

Lab Department

Hematology

CPT Code

85027

LOINC

58410-2 CBC panel Auto (Bld)