Test Code FET Iron, Total
Additional Codes
EPIC | LAB4127 |
Purpose
Alterations in iron and total iron binding capacity levels result from changes in iron intake, absorption, storage, and release mechanisms. Such changes are indicative of a wide range of dysfunctions including anemias, nephrosis, cirrhosis and hepatitis. Both iron and total iron binding capacity measurements are important for definitive diagnosis because they are interrelated.
Source: Beckman Coulter Unicel, FE IFU. REF #467910 12/22/15
Methodology
Timed-endpoint (Beckman Coulter UniCel DxC 660i)
Sample Type
Preferred Sample Type | |
---|---|
Yellow top - Serum |
Acceptable Sample Type(s) | |
---|---|
Red top - Serum | |
Light green top - Plasma |
Centrifuge: Yes
Specimen Minimum Volume
0.3 mL
Specimen Stability
Temperature | Time |
---|---|
Ambient (18-25°C) | 8 hours |
Refrigerated (2-8°C) (store) | 48 hours |
Frozen (-15 to -20°C) | 1 month |
Separate plasma or serum from cells within 2 hours.
Reference Ranges
Normal Range | |
---|---|
0 - 11 months | 100 - 250 ug/dL |
1 - 12 years | 40 - 100 ug/dL |
Adult Male | 45 - 190 ug/dL |
Adult Female* | 30 - 180 ug/dL |
Technical Range | |
---|---|
5 - 500 ug/dL |
*Oral contraceptives may elevate iron values.
Rejection Criteria
Hemolysis | 1+ or greater |
---|---|
Icteric | N/A |
Lipemia | 3+ ultracentrifuge |
EDTA, Sodium citrate, and potassium oxalate specimens.
Availability
Performed | STAT |
---|---|
24/7 | Yes |
Performing Laboratory
MultiCare Yakima Memorial Hospital Laboratory
Lab Department
Chemistry
CPT Code
83540
LOINC
2498-4 Iron [Mass/Vol]