Sign in →

Test Code DBIL Bilirubin, Direct

Important Note

Photosensitive, protect from light following collection!

Additional Codes

EPIC LAB4025

Clinical Significance

Direct bilirubin measurements are used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.

 

Source: Beckman Coulter Unicel, DBIL IFU. REF #439715, #476856 12/22/15

Methodology

Timed endpoint diazo method (Beckman Coulter UniCel DxC 660i)

Sample Type

Preferred Sample Type
Light green top - Plasma

 

Acceptable Sample Type
Dark green top - Plasma
Yellow top - Serum
Red top - Serum

Centrifuge: Yes

Specimen Minimum Volume

0.3 mL

Specimen Stability

Temperature Time
Ambient (18-25°C) 8 hours
Refrigerated (2-8°C) (store) 48 hours
Frozen (-15 to -20°C) 1 month

Separate plasma or serum from cells within 2 hours.

Reference Ranges

Normal Range
0.0 - 0.2 mg/dL

 

Critical Value
>10.0 mg/dL

 

Technical Range
0.1 - 10.0 mg/dL

Rejection Criteria

Hemolysis 2+ or greater reject
Icterus N/A
Lipemia 4+ or greater ultracentrifuge

 

Availability

Performed STAT
24/7 Yes

 

Performing Laboratory

MultiCare Yakima Memorial Hospital Laboratory

Lab Department

Chemistry

CPT Code

82248

LOINC

1968-7 Bilirubin.direct [Mass/Vol]