Test Code DBIL Bilirubin, Direct
Additional Codes
| EPIC | LAB4025 |
Clinical Significance
Measurement of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.
Bilirubin is formed in the reticuloendothelial system during the degradation of aged erythrocytes. The heme portion from hemoglobin and from other heme‑containing proteins is removed, metabolized to bilirubin, and transported as a complex with serum albumin to the liver. In the liver, bilirubin is conjugated with glucuronic acid for solubilization and subsequent transport through the bile duct and elimination via the digestive tract. Diseases or conditions which, through hemolytic processes, produce bilirubin faster than the liver can metabolize it, cause the levels of unconjugated (indirect) bilirubin to increase in the circulation. Liver immaturity and several other diseases in which the bilirubin conjugation mechanism is impaired cause similar elevations of circulating unconjugated bilirubin. Bile duct obstruction or damage to hepatocellular structure causes increases in the levels of both conjugated (direct) and unconjugated (indirect) bilirubin in the circulation.
Source: Roche Cobas BILD2 Method Sheet Version 1.0 2023-12
Methodology
Diazo method (Roche Cobas c503/c303)
Sample Type
| Preferred Sample Type | |
|---|---|
| Light green top - Plasma | |
| Acceptable Sample Type | |
|---|---|
| Dark green top - Plasma | |
| Yellow top - Serum | |
| Red top - Serum | |
Centrifuge: Yes
Specimen Minimum Volume
55 uL with insert cup
Specimen Stability
| Temperature | Time |
|---|---|
| Ambient (15-25°C) | 2 days |
| Refrigerated (2-8°C) (store) | 7 days |
| Frozen (-15 to -25°C) | 6 months |
Separate plasma or serum from cells within 2 hours.
NOTE: Stability limits only valid for light protected specimens.
Reference Ranges
| Normal Range |
|---|
| 0.0 - 0.2 mg/dL |
| Critical Value |
|---|
| >10.0 mg/dL |
| Technical Range |
|---|
| 0.08 - 27.6 mg/dL |
Rejection Criteria
| Hemolysis | Index >25 |
|---|---|
| Icterus | N/A |
| Lipemia | Index >750, ultracentrifuge |
Availability
| Performed | STAT |
|---|---|
| 24/7 | Yes |
Performing Laboratory
MultiCare Yakima Memorial Hospital Laboratory
MultiCare Cancer Care (NSL)
Lab Department
Chemistry
CPT Code
82248
LOINC
1968-7 Bilirubin.direct [Mass/Vol]