Sign in →

Test Code DBIL Bilirubin, Direct

Important Note

Photosensitive, protect from light following collection!

Additional Codes

EPIC LAB4025

Clinical Significance

Measurement of the levels of bilirubin, an organic compound formed during the normal and abnormal destruction of red blood cells, is used in the diagnosis and treatment of liver, hemolytic, hematological, and metabolic disorders, including hepatitis and gall bladder block.  

 

Bilirubin is formed in the reticuloendothelial system during the degradation of aged erythrocytes. The heme portion from hemoglobin and from other heme‑containing proteins is removed, metabolized to bilirubin, and transported as a complex with serum albumin to the liver. In the liver, bilirubin is conjugated with glucuronic acid for solubilization and subsequent transport through the bile duct and elimination via the digestive tract. Diseases or conditions which, through hemolytic processes, produce bilirubin faster than the liver can metabolize it, cause the levels of unconjugated (indirect) bilirubin to increase in the circulation. Liver immaturity and several other diseases in which the bilirubin conjugation mechanism is impaired cause similar elevations of circulating unconjugated bilirubin. Bile duct obstruction or damage to hepatocellular structure causes increases in the levels of both conjugated (direct) and unconjugated (indirect) bilirubin in the circulation. 

 

Source: Roche Cobas BILD2 Method Sheet Version 1.0 2023-12

Methodology

Diazo method (Roche Cobas c503/c303)

Sample Type

Preferred Sample Type
Light green top - Plasma

 

Acceptable Sample Type
Dark green top - Plasma
Yellow top - Serum
Red top - Serum

Centrifuge: Yes

Specimen Minimum Volume

55 uL with insert cup

Specimen Stability

Temperature Time
Ambient (15-25°C) 2 days
Refrigerated (2-8°C) (store) 7 days
Frozen (-15 to -25°C) 6 months

Separate plasma or serum from cells within 2 hours.

NOTE: Stability limits only valid for light protected specimens.

Reference Ranges

Normal Range
0.0 - 0.2 mg/dL

 

Critical Value
>10.0 mg/dL

 

Technical Range
0.08 - 27.6 mg/dL

Rejection Criteria

Hemolysis Index >25
Icterus N/A
Lipemia Index >750, ultracentrifuge

 

Availability

Performed STAT
24/7 Yes

 

Performing Laboratory

MultiCare Yakima Memorial Hospital Laboratory

MultiCare Cancer Care (NSL)

Lab Department

Chemistry

CPT Code

82248

LOINC

1968-7 Bilirubin.direct [Mass/Vol]