Test Code CHGCT Chlamydia, Gonorrhoeae, and Trichomonas PCR
Additional Codes
EPIC | LAB4603 |
Purpose
This assay may be used for detection of Chlamydia trachomatis (CT), Neisseria gonorrhoeae (GC), and/or Trichomonas vaginalis (TV) in urine samples, clinician-collected endocervical swab specimens, and patient-collected vaginal swab specimens (in a clinical setting). The assay is indicated for use to aid in the diagnosis of chlamydial urogenital disease, gonococcal urogenital disease and/or trichomoniasis in asymptomatic and symptomatic individuals.
Methodology
The BD MAX CT/GC/TV assay, as performed using the BD MAX system, incorporates automated DNA extraction and real-time polymerase chain reaction (PCR) for the direct, qualitative detection of DNA from Chlamydia trachomatis (CT), Neisseria gonorrhoeae (GC), and/or Trichomonas vaginalis (TV).
Sample Type
- Urine
- Vaginal/Endocervical Swab
Specimen Minimum Volume
2 mL
Specimen Stability
Specimen Stability Prior to Transfer into the BD MAX UVE Sample Buffer Tube
Specimen Stability | Specimen Type | Transport/Storage Temperature | |
---|---|---|---|
2 - 30°C | 2 - 8°C | ||
Prior to aliquoting into Sample Buffer Tube | Urine | 4 hours | 24 hours |
Vaginal/Endocervical Swab | 2 hours | 2 hours |
Sample Stability in BD MAX UVE Sample Buffer Tube Prior to Pre-Warm
Specimen Stability | Specimen Type | Transport/Storage Temperature | |
---|---|---|---|
2 - 30°C | - 20°C | ||
In Sample Buffer Tube | Urine | 5 days | 30 days |
Vaginal/Endocervical Swab | 5 days | 30 days |
Specimen Collection
Urine
- Uncap the Sample Buffer Tube (SBT). Use the graduated transfer pipette to mix the urine specimen gently in the collection cup and transfer approximately 2 mL of the urine into the SBT.
- Tighten the cap securely on the SBT.
- Invert the SBT 3-4 times to ensure specimen and reagent are well mixed.
- Label the SBT with patient identification, source, and date/time collected. Do not obscure the bar codes on the tube!
- Refrigerate if testing will be within 5 days.
Endocervical or Vaginal Swab
- Uncap the Sample Buffer Tube (SBT) and fully insert the swab into the tube so that the tip is at the botom.
- Carefully break the swab shaft at the score mark. Use caution to avoid splashing or contamination of the tube contents.
- Tighten the cap securely on the SBT.
- Label the SBT with patient identification, source, and date/time collected. Do not obscure bar codes on the tube.
- Refrigerate if testing will be within 5 days.
Rejection Criteria
- Samples outside of stability limits.
Availability
Performed | STAT |
---|---|
Daily at 1000, 1400, and 1900 | No |
Performing Laboratory
MultiCare Yakima Memorial Hospital Laboratory
Lab Department
Microbiology
CPT Code
87491
87661
87591
LOINC
92684-0 Chlamydia trachomatis and Neisseria gonorrhoeae and Trichomonas vaginalis DNA panel NAA+probe (U)